PAP-Mediated Alveolar Recruitment Demonstrated by Serial Arterial Blood Gas Analysis During Attended Full-Night BIPAP Titration in a Patient with HFpEF, Morbid Obesity, and AKI on CKD
A case study demonstrating how serial intranight ABG analysis during BiPAP titration confirmed EPAP-driven alveolar recruitment and shunt reversal in a patient with HFpEF, morbid obesity, and AKI on CKD.
A case study demonstrating how serial intranight ABG analysis during BiPAP titration confirmed EPAP-driven alveolar recruitment and shunt reversal in a patient with HFpEF, morbid obesity, and AKI on CKD.

Updated:
22 October 2026
ABSTRACT
Background: Serial arterial blood gas (ABG) sampling during attended polysomnography with positive airway pressure (PAP) titration provides a physiological window into overnight gas exchange dynamics. We report the use of four-point intranight ABG analysis during full-night BiPAP Volume-Assured Pressure Support (VAPS) titration in a complex multimorbid patient.
Case: A 46-year-old male with Heart Failure with Preserved Ejection Fraction (HFpEF), Acute Kidney Injury on Chronic Kidney Disease (AKI on CKD), metabolic syndrome, morbid obesity (BMI 31.5 kg/m²), and obstructive sleep apnea underwent attended full-night PAP titration polysomnography. Serial ABGs were obtained at four timepoints: pre-study baseline (FiO₂ 21%), evening decline, mid-study on BiPAP alone (room air), and morning on BiPAP plus 1L/min O₂.
Findings: Evening pulmonary shunt fraction (FShunt) rose to 14.1% with pO₂ 60.6 mmHg on room air, consistent with obesity-related positional atelectasis and HFpEF fluid redistribution. After BiPAP titration (IPAP 12/EPAP 6 cmH₂O, VAPS, TV 500 mL, RR 14/min, Ti 1.3s) on room air alone, FShunt reduced to 5.6% with pO₂ 78.9 mmHg — demonstrating EPAP-driven alveolar recruitment without supplemental oxygen. Addition of 1L/min O₂ produced supraphysiologic oxygenation (pO₂ 157 mmHg, FShunt -9.6%). Acid-base parameters remained stable throughout (compensated non-anion-gap metabolic acidosis; pH 7.33–7.37, pCO₂ 35–38.5 mmHg), confirming adequate ventilatory support via the VAPS backup mechanism. Overall AHI was 2.0/h on optimal settings. No Cheyne-Stokes respiration was detected.
Conclusion: This case demonstrates that EPAP-mediated alveolar recruitment, independently of supplemental oxygen, can reverse significant positional shunt in patients with obesity and HFpEF during BiPAP titration. Serial intranight ABG correlation provides mechanistic insights beyond routine PSG indices and identifies the dominant pathophysiological mechanism — recruitable shunt versus fixed parenchymal disease — which has direct therapeutic implications.
Keywords: Arterial blood gas; BiPAP titration; shunt fraction; alveolar recruitment; obesity hypoventilation; HFpEF; polysomnography; VAPS; EPAP
1. INTRODUCTION
Routine attended polysomnography with positive airway pressure (PAP) titration is the gold-standard diagnostic and therapeutic procedure for sleep-disordered breathing. Standard PSG scoring yields data on apnoea-hypopnoea index (AHI), arousal index, sleep architecture, and pulse oximetry trends. However, these parameters provide limited mechanistic insight into the gas exchange physiology driving nocturnal hypoxaemia — particularly in patients with multiple overlapping cardiorespiratory comorbidities.
Arterial blood gas (ABG) analysis directly measures the partial pressures of oxygen and carbon dioxide, acid-base status, and derived indices including pulmonary shunt fraction (FShunt) and alveolar-arterial (A-a) oxygen gradient. The ABL800 FLEX (Radiometer) analyser calculates FShunt from two-point ABG data, enabling real-time estimation of intrapulmonary shunt — the proportion of cardiac output perfusing unventilated alveoli.
In patients with morbid obesity and cardiac failure, multiple simultaneous mechanisms can produce nocturnal hypoxaemia: obstructive sleep apnoea, positional atelectasis with derecruitment, pulmonary congestion from cardiac fluid shifts, and hypoventilatory CO₂ retention. These mechanisms respond differently to therapeutic interventions — alveolar recruitment responds to EPAP; pulmonary oedema requires diuresis; hypoventilation responds to volume-assured ventilation. Distinguishing between these mechanisms in real-time has direct implications for PAP titration decisions.
We report this unique case, in which four-point serial ABG analysis was performed in temporal correlation with attended BiPAP VAPS titration, demonstrating dynamic shunt reversal by EPAP-mediated alveolar recruitment in a patient with HFpEF, AKI on CKD, and morbid obesity.
2. CASE PRESENTATION
2.1 Patient Profile
A 46-year-old male presented to Sri Aurobindo Medical College & PGI, Indore, under Ayushman Bharat Yojana (insurance-funded care) admitted to Medicine Ward. Presenting complaints included breathlessness on exertion, leg swelling (?cellulitis, right lower limb), and excessive daytime sleepiness.
Comorbidity Profile:
Heart Failure with Preserved Ejection Fraction (HFpEF) — EF preserved on prior echo
Acute Kidney Injury on Chronic Kidney Disease (AKI on CKD)
Metabolic Syndrome (central obesity, hypertension, dysglycaemia, dyslipidaemia)
Morbid Obesity — BMI 31.5 kg/m², Height 170 cm, Weight 91 kg
Non-Proliferative Diabetic Retinopathy
Cellulitis — Right lower limb
Obstructive Sleep Apnea (clinically suspected, confirmed on PSG)
Haematological and biochemical context: Haemoglobin 7.9–8.4 g/dL (mild-moderate anaemia of CKD); persistent hypocalcaemia (ionised Ca²⁺ 1.07–1.14 mmol/L); mild hyponatraemia on admission (Na 134 mmol/L). Blood glucose was controlled across the study (81–95 mg/dL).
2.2 Pre-Titration Assessment
Clinical assessment at 01:13 AM on 12 June 2026 (ABG-1, room air) confirmed a baseline of: pH 7.337, pCO₂ 36.9 mmHg, pO₂ 102 mmHg, HCO₃⁻ 19.2 mmol/L, FShunt -2.1%. This established a compensated non-anion-gap (hyperchloremic) metabolic acidosis, consistent with AKI on CKD, with near-zero pulmonary shunt at baseline.
Winter's formula expected pCO₂ for HCO₃⁻ ~20 mmol/L = 36–40 mmHg; measured pCO₂ 36.9 confirms appropriate respiratory compensation without superimposed primary respiratory disorder at baseline.
By the next evening (21:36 hrs, ABG-2, room air), oxygenation had deteriorated significantly: pO₂ fell to 60.6 mmHg, FShunt rose to 14.1%, A-a gradient 37.9 mmHg, pO₂/FiO₂ ratio 288 mmHg. Acid-base remained stable (pH 7.366, pCO₂ 35.6 mmHg). This dissociation — oxygenation deterioration with stable ventilation — pointed to a V/Q mismatch or shunt mechanism rather than primary hypoventilation.
2.3 Polysomnography Procedure
Full-night attended PAP titration polysomnography (Level 1,) was performed on 12–13 June 2026 using the Sleepware G3 platform (Philips Respironics). Scoring per AASM criteria (30-second epochs); hypopnoeas scored with 3% desaturation criterion (AASM ). Parameters monitored: frontal/central/occipital EEG, bilateral EOG, submentalis EMG, nasal/oral airflow, thoracoabdominal inductance plethysmography, ECG, pulse oximetry (SpO₂), body position, snore channel. Acquiring Technician: Makhan Singh Boriya.
BiPAP VAPS S/T mode (Philips OMNILAB) was initiated with: Tidal Volume target 500 mL, Backup RR 14/min, Ti 1.3 seconds. IPAP/EPAP titration commenced at 8/4 cmH₂O and was escalated per standard attended titration protocol.
2.4 Sleep Architecture Findings
Parameter | Value | Normal | Interpretation |
Total Sleep Time | 302.0 min | >360 min | Reduced |
Sleep Efficiency | 69.2% | >85% | Abnormal ↓ |
WASO | 124.7 min | <30 min | Markedly elevated |
N1 Sleep | 15.5 min (5.1%) | 2–5% | Normal |
N2 Sleep | 222.0 min (73.5%) | 45–55% | Dominant |
N3 Slow Wave Sleep | 47.0 min (15.6%) | 15–20% | Low normal |
REM Sleep | 17.5 min (5.8%) | 20–25% | Severely suppressed |
REM Latency | 114.1 min | <120 min | Borderline delayed |
Leg Movement Arousal Index | 3.8/h | <5/h | Borderline — monitor |
REM suppression to 5.8% of TST is a recognised pattern in HFpEF and obesity-related sleep-disordered breathing, where REM instability leads to protective REM curtailment. The absence of Cheyne-Stokes respiration throughout is a notable finding in the context of HFpEF.
2.5 Respiratory Findings on PAP Titration
Overall AHI was 2.0/h on the optimal settings, representing excellent respiratory control. Two obstructive apnoeas and seven hypopnoeas were recorded across the full night, predominantly in NREM. No REM respiratory events were recorded (REM AHI 0.0/h). No Cheyne-Stokes Breathing was detected.
Pressure distribution analysis identified IPAP 12/EPAP 6 as the optimal setting (156.6 min TIB, AHI 2.0/h, minimum SpO₂ 93%). Settings with EPAP 5 consistently showed inferior SpO₂ performance (minimum 85% at IPAP 12/EPAP 5) or higher AHI. EPAP 6 was the minimum effective EPAP threshold for this patient.
2.6 The Key Finding — Serial ABG Trajectory
Four arterial blood gas analyses were performed on the Radiometer ABL800 FLEX analyser (ABG-S 195µL arterial samples, corrected to 37°C, FiO₂ 21% for timepoints 1–3; approximately 24% FiO₂ for timepoint 4 on 1L/min O₂):
Parameter | ABG-1 01:13 AM RA, Pre-study | ABG-2 21:36 PM RA, Evening | ABG-3 03:45 AM BiPAP, RA | ABG-4 07:24 AM BiPAP+1L O₂ |
pH | 7.337 | 7.366 | 7.341 | 7.331 |
pCO₂ (mmHg) | 36.9 | 35.6 | 37.0 | 38.5 |
pO₂ (mmHg) | 102 | 60.6 ↓ | 78.9 | 157 ↑ |
HCO₃⁻ (mmol/L) | 19.2 | 19.9 | 20.0 | 19.8 |
Base Excess | -5.5 | -4.5 | -5.3 | -5.1 |
FShunt (%) | -2.1 | 14.1 ↑ | 5.6 | -9.6 ↓ |
A-a Gradient (mmHg) | — | 37.9 | 18.0 | — |
pO₂/FiO₂ (mmHg) | 485 | 288 | 376 | 749 |
Setting | None | None | IPAP12/EPAP6 | IPAP12/EPAP6+O₂ |
RA = Room Air. FShunt calculated by ABL800 FLEX from measured and alveolar gas values.
3. DISCUSSION
3.1 Mechanism of Shunt Rise — Evening Derecruitment
The evening rise in FShunt from -2.1% to 14.1% (ABG-1 to ABG-2) in the absence of any acid-base change or CO₂ rise points to a pure oxygenation deficit — mechanistically consistent with alveolar derecruitment. In a patient with morbid obesity (BMI 31.5) and HFpEF, the supine posture and fluid redistribution from dependent periphery to central circulation in the evening creates the ideal conditions for basilar atelectasis: increased abdominal pressure restricts diaphragmatic excursion, while pulmonary venous engorgement from HFpEF reduces alveolar compliance. Neither mechanism involves primary ventilatory failure; hence pCO₂ and acid-base are preserved.
The A-a gradient of 37.9 mmHg at this point (normal <10–15 mmHg on room air) with an intact ventilatory response (pCO₂ 35.6 mmHg) is the classic fingerprint of V/Q mismatch or shunt, not dead space-predominant physiology.
3.2 EPAP-Mediated Recruitment — The Therapeutic Mechanism
The most physiologically significant finding is the reduction of FShunt from 14.1% to 5.6% on BiPAP alone, room air, between ABG-2 and ABG-3 — representing approximately 60% shunt reduction purely through EPAP-maintained alveolar patency.
EPAP in BiPAP functions analogously to PEEP in invasive ventilation — it provides a continuous end-expiratory pneumatic splint that prevents alveolar collapse at end-expiration. In an obese patient with reduced functional residual capacity (FRC), an EPAP of 6 cmH₂O provided sufficient transmural pressure to maintain basilar alveolar patency throughout the expiratory phase, reversing the derecruitment that had accumulated through the waking hours.
This mechanism is distinct from the obstructive apnoea-resolution effect of EPAP (upper airway pneumatic stenting) and from the ventilatory CO₂ clearance effect of IPAP-EPAP pressure support. It represents a pure parenchymal-level gas exchange improvement, independent of respiratory event termination.
3.3 VAPS and Acid-Base Stability
The stability of acid-base parameters across all four ABG timepoints is equally important. pH (7.331–7.366), pCO₂ (35.6–38.5 mmHg), and HCO₃⁻ (19.2–20.0 mmol/L) remained essentially constant across a 6-hour period spanning no treatment, BiPAP initiation, and O₂ addition. This confirms two things: (i) the underlying metabolic acidosis (non-anion-gap, hyperchloremic, likely renal tubular acidosis from AKI on CKD) was not acutely exacerbated by the study procedure; and (ii) the BiPAP VAPS S/T with backup RR 14/min and Ti 1.3s maintained adequate minute ventilation throughout, preventing both CO₂ retention and overcorrection alkalosis.
Winter's formula (expected pCO₂ = 1.5 × HCO₃ + 8 ± 2 = 36–40 mmHg for HCO₃ ~20 mmol/L) was met at every timepoint, confirming that this patient's respiratory compensation for the metabolic acidosis remained intact throughout and was not disturbed by the ventilatory support delivered.
3.4 Absence of Cheyne-Stokes Respiration
The absence of Cheyne-Stokes respiration (CSR) in a patient with HFpEF is noteworthy. CSR in cardiac failure arises from prolonged circulatory delay causing oscillation between hyperpnoea and apnoea, and is typically seen with reduced ejection fraction, elevated LV filling pressures, and blunted central chemoreceptor gain. Its absence here may reflect: preserved EF (HFpEF rather than HFrEF), adequate central respiratory drive (consistent with normal resting pCO₂), or effective fluid redistribution management. This finding also confirms that the hypoxaemia in this case was peripheral (parenchymal/atelectatic) rather than central in origin.
3.5 Clinical Implications
This case highlights the limitations of pulse oximetry-only monitoring during attended PAP titration. Pulse oximetry would have identified the evening hypoxaemia and prompted empirical O₂ supplementation — but would not have distinguished between atelectatic shunt (responsive to EPAP) and pulmonary oedema (requiring diuresis) or intrinsic parenchymal shunt (potentially fixed). The ABG data allowed a mechanistically informed decision: EPAP titration, not FiO₂ augmentation.
In complex patients with overlapping cardiorespiratory comorbidities — HFpEF, CKD, morbid obesity, metabolic syndrome — routine PSG titration without ABG correlation risks optimising for AHI while leaving the dominant gas exchange pathology (shunt, V/Q mismatch, hypoventilation) unaddressed or misidentified.
4. CONCLUSION
We report the dynamic, real-time reversal of pulmonary shunt during attended BiPAP VAPS titration, demonstrated through four-point serial arterial blood gas analysis. EPAP-mediated alveolar recruitment, independent of supplemental oxygen or respiratory event termination, reduced FShunt from 14.1% to 5.6% — establishing recruitable atelectasis as the operative mechanism of hypoxaemia in this complex patient. VAPS ventilatory support maintained acid-base balance throughout the study.
Serial intranight ABG correlation during attended PSG titration is a feasible and informative protocol in Level 1 sleep laboratories equipped for arterial sampling. It provides mechanistic granularity beyond AHI and SpO₂ trends, enabling physiologically targeted PAP prescription in patients with multiple cardiorespiratory comorbidities. Wider adoption of this protocol in complex inpatient sleep studies is recommended.
5. PATIENT CONSENT
Informed written consent was obtained from the patient for clinical management, publication of deidentified case data, and photographic/laboratory documentation. No identifying personal information (name, address) has been included in this report per institutional anonymisation guidelines.
6. DECLARATIONS
Conflict of Interest: The author holds advisory roles with Deckmount Electronics and SlumberSense Global. These organisations did not fund or influence this case report. The BiPAP titration was conducted using clinical equipment available at SAMC & PGI.
Funding: None.
Author Contributions:
Dr. Razzakur Rahman: Study conception, patient care, PSG acquisition and scoring, ABG correlation, interpretation, and manuscript preparation.
Dr Jaishree Tapadia: Study conception, Manuscript preparation
Makhan Singh Boriya : Patient care and PSG acquisition, ABG samples
Ethical Statement: The study was conducted in accordance with the Declaration of Helsinki and institutional ethical standards for case reporting.
7. KEY REFERENCES
1. Berry RB, et al. AASM Manual for Scoring of Sleep and Associated Events, Version 2.6. American Academy of Sleep Medicine, 2020.
2. Masa JF, et al. Obesity hypoventilation syndrome. Eur Respir Rev. 2019;28(151):180097.
3. Mokhlesi B, et al. Obesity hypoventilation syndrome: prevalence and predictors in patients with obstructive sleep apnea. Sleep Breath. 2007;11(2):117–124.
4. Oldham MA, Bhatt DL, Bhatt P, et al. Overnight fluid shifts and their relationship to sleep apnea and heart failure. J Clin Sleep Med. 2021.
5. Ranieri VM, et al. Physiological effects of positive end-expiratory pressure in patients with acute lung injury. Am J Respir Crit Care Med. 1993.
6. Simonsson BG, Skoogh BE, Bergh NP. Arterial blood gas estimation using shunt fraction analysis. Acta Anaesthesiol Scand. 1997.
7. Radovanovic D, et al. Arterial blood gas monitoring during non-invasive ventilation: clinical utility in complex sleep disorders. J Sleep Res. 2020.





Our Authors

Get the Latest Healthcare
Stories in Your Inbox.
Subscribe to our newsletter to receive new articles, case insights, and thought leadership from clinicians and researchers worldwide.
.png)